High Accuracy, High Performance and Reliable Measurements
• 0.02% basic DC accuracy
• Large display with 50,000 counts
• Dual display for double indication for AC and DC, V and Hz, etc.
• True-RMS Measurements AC and AC+DC
• Wide AC Frequency bandwidth from 10Hz to 100kHz
• True-RMS or MEAN value detection mode can be selected
• Fast Peak Hold response time of 250µs
• Low-pass filter for motor drive measurements
• Low Power Ω for resistance measurements on sensible electronic circuits by a low and safe test current
• User calibration function

Safety design for industrial use
• Complies with IEC 61010-1 CAT IV 600V, CAT III 1000V
• Terminal shutter to prevent incorrect test leads' insertion in current terminals
• Very wide operating temperature range from -20 to 55ºC

Reliable support for data management
• Large data internal memory
KEW 1062: 10,000 data (logging), 100 data (manual)
• Download data and Live Monitoring on a PC via the USB interface
(Option for USB Communication set)
* MAX 32,767 data can be transferred.

-Detection mode: MEAN/RMS (switch)

-DC V: 50.000/500.00/2400.0mV/5.0000/50.000/500.00/1000.0V
(Input impedance: Approx. 100MΩ [50/500/2400mV], 10MΩ [5/50/500/1000V])
±0.02%rdg±2dgt (Basic accuracy)

-AC V [RMS]: 50.000/500.00mV/5.0000/50.000/500.00/1000.0V
(Input impedance: 11MΩ<50pF [50/500mV/5V], 10MΩ<50pF [50/500/1000V])
±0.4%rdg±30dgt (Basic accuracy)

-AC V [MEAN]: 50.000/500.00mV/ 5.0000/50.000/500.00/1000.0V
(Input impedance:11MΩ<50pF [50/500mV/5V],10MΩ<50pF [50/500/1000V])
±1%rdg±30dgt (Basic accuracy)

-DCV+ACV: 5.0000/50.000/500.00/1000.0V
(Input impedance: 11MΩ<50pF [5V], 10MΩ<50pF [50/500/1000V])
±0.5%rdg±10dgt (Basic accuracy)

-DC A: 500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A
±0.2%rdg±5dgt (Basic accuracy)

-AC A [RMS]: 500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A
±0.75%rdg±20dgt (Basic accuracy)

-AC A [MEAN]: 500.00/5000.0µA/50.000/500.00mA/ 5.0000/10.000A
±1.5%rdg±20dgt (Basic accuracy)

-DCA+ACA: 500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A
±1%rdg±10dgt (Basic accuracy)

-Ω: 500.00Ω/5.0000/50.000/500.00kΩ/5.0000/50.000MΩ
±0.05%rdg±2dgt (Basic accuracy)

-LowPower-Ω: 5.000/50.00/500.0kΩ/5.000MΩ
±0.2%rdg±3dgt (Basic accuracy)

-Continuity buzzer: 500.0Ω (The buzzer turns on for resistances lower than 100±50Ω)

-Diode test: 2.4000V ±1%rdg±2dgt
Open curcuit voltage: <5V (Approx. 0.5mA Measuring Current)

-Capacitance: 5.000/50.00/500.0nF/5.000/50.00/500.0µF/5.000/50.00mF
±1%rdg±5dgt (Basic accuracy)

-Frequency: 2.000 - 9.999/9.00 - 99.99/90.0 - 999.9Hz/0.900 - 9.999/9.00 - 99.99kHz
±0.02%rdg±1dgt (Basic accuracy)

-DUTY: 10 - 90% ±1%rdg

-Temperature: -200 - 1372ºC ±1%rdg±1.5ºC
-328 - 2501.6ºF ±1%rdg±2.0ºF
(with the use of K-type Temperature probe)

-Measurement functions: DC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature, Capacitor, Duty cycle ratio, Decibel (dBV, dBm), Continuity Check, Diode Test, LowPower-Ω

-Other functions: Data Hold (D/H), Auto Hold (A/H), Peak Hold (P/H), Range Hold (R/H), Maximum value (MAX), Minimum value (MIN), Average value (AVG), Zero Adjustment (Capacitor, Resistance), Relative values, Save to Memory, LCD backlight

-Applicable standards: IEC 61010-1 CAT IV 600V, CAT III 1000V Pollution degree 2,
IEC 61010-031, IEC 61326-1 (EMC)

-Power source: R6 (1.5V)×4 (Auto power off: approx. 20 minutes)

-Dimensions: 192(L)×90(W)×49(D) mm

-Weight: Approx. 560g (including batteries)

-Included Accessories: 7220A (Test leads), R6×4, Instruction manual,
8926 (Fuse [440mA/1000V])×1 (included)
8927 (Fuse [10A/1000V])×1 (included)

Ý kiến của bạn